Compact Measurement for Carbon Dioxide and Oxygen -- CarbonOxy

Summary

Features

Applications

Technical Specifications


Summary

The CarbonOxy is a unique device for measuring CO2 and O2 on a Euro Printed Circuit Board (PCB). Carbon Dioxide is measured using the dual wavelength technique (NDIR - Non-Dispersive Infra Red) while Oxygen is measured using Zirconium Oxide. This combination ensures a quick response time along with elimination of intereference from other gases.

With various DIP switches on the circuit board, the response time for the analyzer can be adjusted between 1 to 10 seconds. It is possible to do automatic temperature correction or zero point adjustment can be made. The entire circuit board is constantly monitored by a microcontroller and LEDs in various colors can indicate alarm states and out of range values.

The power consumption of the combination is maximum of 6 watts and hence is suitable for a portable, compact measuring device. All electronics meet the latest European EMV regulations.

>>Top


Features

  • Specific to CO2 and O2 -- no interference from other gases
  • No reference gases are needed
  • Self-calibrating
  • Long lifespan
  • Stable operations
  • Low power consumption
  • Compact device

>>Top


Applications

  • Horticulture green houses
  • Fruit storage
  • Mushroom farming
  • Food packaging
  • Breweries
  • Incubators
  • Livestock Husbandry
  • Landfill gas monitoring

>>Top


Technical Specifications

CO2 O2 CO2/O2
Principle of Operation NDIR Zirconia Oxide NDIR / Zirconia Oxide
Accuracy +/- 2% Full-scale +/- 0.5% Full-scale +/- 2% CO2 and +/- 0.5%O2 of Full-scale
Repeatability +/- 1% Full-scale +/- 1% Full-scale +/- 1% Full-scale for both
Response time T90 1..10 sec (adjustable) 1..10 sec (adjustable) 1..10 sec (adjustable)
Operating Temperature -10 to +50degC -10 to +50degC -10 to +50degC
Zero Drift 0 0 0
Warm-up Time 2 min 2 min 2 min
Humidity 98% non-condensing 98% non-condensing 98% non-condensing
Power Supply 12 to 30 VDC 12 to 30 VDC 12 to 30 VDC
Power Consumption 3W 3W 6W
Outputs 0/4...20mA
0/2...10VDC
RS232
0/4...20mA
0/2...10VDC
RS232
2 x 0/4...20mA
2 x 0/2...10VDC
RS232
Lfe span > 3 years > 3 years > 3 years
Flow rate 0.5 l/min 0.5 l/min 0.5 l/min
Dimensions 160 x 100 x 40 mm
6.3 x  3.9 x 1.6 in
160 x 100 x 40 mm
6.3 x  3.9 x 1.6 in
160 x 100 x 40 mm
6.3 x  3.9 x 1.6 in
Weight 300 gm (0.66 lb) 280 gm (0.62lb) 380 gm (0.84lb)
Measurement Ranges O2 CO2
0 to 5% na x
0 to 10% na x
0 to 25% x x
0 to 50% x x
0 to 95% x na
0 to 100% na x

x -- available        na-- not available