Compact Real-Time Photoelectric Aerosol Sensor
(PAS 2000CE)
-
Detects ultrafine carbon aerosols in
real-time
-
Lightweight, compact and ruggedized
construction
-
Battery powered with onboard data storage
-
Applications related to personal exposure in
various environments (homes, workplace, automobiles, tunnels etc.)
-
Companion product to the DC
2000CE --- the DC 2000CE measures ALL particulate matter, the PAS 2000CE
measures PAH (primarily on carbonaeous matter)
>>Top
The PAS 2000 works on the following principles:
• Using an Excimer lamp the aerosol flow is exposed to UV radiation.
The Excimer lamp offers a high intensity, narrowband source of UV
radiation. The wavelength of the light is chosen such that only the
carbon aerosols are ionized, while gas molecules and non-carbon
aerosols remain neutral. • The carbon aerosol particles emit
electrons, which are subsequently removed when an electric field is
applied. • The remaining positively charged particles are
collected on a filter inside an electrometer, where the charge is
measured. The resulting electric current establishes a signal, which
is proportional to the concentration of Elemental Carbon and/ or PAH.
Also by operating the Excimer lamp in a chopped mode, the PAS 2000
can eliminate the background signal, which is associated with freshly
generated aerosols. The chopped mode also enables a dynamic zero to be
generated automatically by the instrument.>>Top
Source-specific calibration curves are
available or can be generated where the monitor output is compared to an
analytically determined EC and PAH concentration. A site-specific calibration
curve can provide greater accuracy for the particle size, charge and
compound-specific to the source. In addition to the site-specific curves, an
approximate universal calibration curve can be used for screening and real-time
trending applications.
>>Top
|
Display
|
16
characters with 2 lines LED
|
|
Power
|
115
volts AC / 60 Hz & 220
volts AC / 50 HzMax.
Battery
15 volts Lithium Metal Hydride
|
|
Range
|
0
to 4000 ng / m3
|
|
Lower
Threshold
|
~
1 ng / m3
|
|
Response
time
|
<
10 seconds (adjustable)
|
|
Digital
Output
Analog
Output
|
RS
- 232 (for data download and program upload)
0
to 10 volts
|
|
Sample
gas
|
Built-in
pump with flowrate controlled at 1 L/min
|
|
Operating
temp
|
40
to 104 ºF (5 to 40ºC)
|
|
Dimensions
|
Height
x Width x Depth = 3in x 7in x 5in ( 68mm x 175mm x 124mm )
|
|
Weight
|
3
lb. ( 1.5 kg )
|
|
Data
Storage
Data
Download
|
8000
data points (each
data point consisting of : Time, Value )
User-friendly
PC-compatible graphical software used for downloading the collected data.
Flat ASCII file output can also be generated for further analysis (e.g.
Microsoft Excel format).
|
>>Top
|