n

 EcoChem Analytics ... Always the Best That We Can Be 

  202 Reynolds
  League City, TX 77573, USA
  Tel: 281-338-9888  Fax: 281-332-6152
  Email: info@ecochem.biz

.. 

n

Home

Products
    Stack Monitors
    Mercury
    PAH  
    Particulates
    Gas Analyzers
    Hovacal
Brochures
Pricing and Quotations
Installations
Related Sites
Photo Gallery
Library
Site Map
Users' Area
Contact

HG-MONITOR 3000

Summary
Measuring Principle
Functionality
UV Light Source
Easy Operation
Fields of Application
Data Display and Output
Self-Diagnostics System
Technical Specifications
Partner

>>Top


Summary

  • Continuous real-time online operation
  • Easy to operate, self-explaining software
  • Auto-zero for precise measurement
  • Available for tabletop, 19" rackmount and mobile use
  • Measurement range of 0.1 to 3000 µg/m3 (non-linear compensation in high measuring range )
  • Temperature control for light-source and cuvette
  • Unique electrodeless mercury lamp with long lifetime (over 20,000 hours)
  • Wide range power supply (90 - 265 VAC/VDC, 12 VDC)
  • Graphic TTF-display and one-touch keypad
  • External Keyboard and VGA monitor, remote control by modem possible
  • Approved for EMC and low-voltage directives (CE)
  • Main component ("heart") of a complete range of mercury analysers for gases, liquids, solids
  • RS 232, RS485, 4-20 mA analog and PLC outputs
  • Self-diagnosis, status outputs, printer port

>>Top


Measuring Principle

The measuring principle used in the HG - MONITOR 3000 is based on the resonance absorption of mercury atoms at a wavelength of 253.7 nm. It exploits the high vapor pressure of mercury and the fact that except for the noble gases, Mercury is the only element that has monoatomic vapor at room temperature.

>>Top


Functionality

The sample gas is continuously fed via a dust filter into the optical cuvette of the HG - MONITOR 3000. Here an absorption measurement of the radiation emitted by a UV light source (253.7 nm Hg line) is carried out. The intensity of the UV lamp is continuously measured via a reference beam and the measurement is corrected accordingly to compensate for possible fluctuations of the light source.

>>Top


UV Light Source

The HG - MONITOR 3000's UV source is a low-pressure mercury discharge lamp with high-frequency excitation. This method yields extreme-narrow-band emission lines, which are congruent with the absorption lines of the mercury atoms. This provides improved detection sensitivity and a lower cross-sensitivity compared to other light sources. The lamp has a much longer service life than other UV lamps. The lamp unit is temperature-stabilized.

>>Top


Easy Operation

The HG - MONITOR 3000 is micro-processor-controlled. After it is switched on the device automatically calibrates itself and then switches to measuring mode. Inputs can be made via the waterproof membrane keyboard. The following parameters are menu-adjustable:

  • Display of ug/m3 or ppb
  • Conversion to standard conditions
  • Time interval for printer output
  • Alarm level and prelim level
  • 3 time intervals for computation of mean value
  • Measuring range 0...100, 0...1000, 0...2000 ug/m3 (ranges are user programmable)

>>Top


Fields of Application

  • Mercury in air
  • Mercury in gases such as H2 , N2 and Natural Gas.
  • Work-place monitoring
  • TLV (Threshold Limit Value) monitoring
  • Measurement of mercury in exhaust air
  • Ambient air screening at ground level
  • Detector for amalgam technology

>>Top


Data Display and Output

The mercury concentration is displayed on a 4-line display and is also outputted to a printer interface and an RS- 232 interface. The output is adjusted for pressure and temperature. Mean values over programmable periods of time (e.g. half-hour mean values) can also be calculated. The pressure and temperature of the sample gas can be displayed by pressing a button. An analog output signal is also available for connecting to a data acquisition system or a chart recorder. 

>>Top


Self-Diagnostics system

The built-in processor constantly monitors the functions of vital components of the HG - MONITOR 3000. If a fault occurs the user is warned via the display and via a separate output signal.

>>Top


Technical Specifications

Measuring Method

UV Absorption

Wavelength

254 nm

UV Lamp

Temperature-controlled high frequency excitation lamp

Cuvette

240 mm, quartz glass

Measuring Ranges

0-100, 0-1000, 0-3000 ug/m3 (ranges are user programmable)

(Gaseous phase)

0-25, 0-125, 0-250 ppb

Accuracy

± 2 % at 100 ug/m3

Sensitivity

0.1 ug/m3

Temp/Pressure Compensation

Automatic

Response Time

Selectable, 0 / 1 / 5 Seconds

Calibration

Factory-calibrated

Alarm

2 levels programmable, over full scale

Mean Value

Selectable between 0-999 min

Output Signals

Analog : 0-1 V (optionally 4...20 mA) Digital : Centronics and RS 232~ Status : (Alarm and Error Message)

Display

4-line LCD with background illumination

Pump

Membrane pump, ~ 1.5 l/min

Filter

Dust filter, 50 mm diameter

Power Supply

110 / 127 / 220 / 240 VAC (option: 12 V operation from rechargeable battery), 50 / 60 Hz

Operating Temperature

32 to 113 °F ( 0 to 45 °C )

Dimensions

(Width x Height x Depth) without handle  
18 in x 5.2 in x 13 in (449 mm x 133 mm x 331 mm)  
Optional version: 19" chassis unit

Weight

16.5 lb. ( 7.5 kg )

>>Top


Partner

The HG-MONITOR 3000 is made by Seefelder Messtechnik in Germany.

>>Top


 

 

 

......

Home Stack Monitors Mercury PAH Particulates Gas Analyzers Hovacal Brochures
Pricing Installations Related Sites Photo Gallery Library Site Map Users'Area Contact

copyright 2001-2005 EcoChem Analytics